Linear relationship between cytoplasm resistance and hemoglobin in red blood cell hemolysis by electrical impedance spectroscopy & eight-parameter equivalent circuit
Linear relationship between cytoplasm resistance and hemoglobin in red blood cell hemolysis by electrical impedance spectroscopy & eight-parameter equivalent circuit
复制标题
电阻抗光谱法研究红细胞溶血中细胞质电阻与血红蛋白的线性关系
DOI:
10.1016/j.bios.2018.08.012
复制
发表时间:
2018
影响因子:
12.6
通讯作者:
Takei Masahiro
中科院分区:
文献类型:
--
作者:
Tran A. Kiet;Sapkota Achyut;Wen Jianming;Li Jianping;Takei Masahiro
We have developed a non-invasive rapid and real-time red blood cell (RBC) hemolysis detection method which is a more accurate for point of care testing of hemolysis in various medical settings. An eight-parameter equivalent circuit is employed to quantify the release of hemoglobin (Hb) and the cytoplasm from RBC into the blood plasma. RBC hemolysis is induced by adding different volume fractions of distilled water into the blood. The cytoplasm released following RBC hemolysis is estimated from the experimental values. A strong relationship between RBC hemolysis and change in the electrical characteristics of blood has been demonstrated. The cytoplasm resistance (Rc) shows a linear relationship with theHb. This relationship betweenRcandHbis described by the equationRc= 0.2203Hb+ 2.4775, with a correlation coefficient of 0.9905.