Detection efficiency of a heavy ion time‐of‐flight spectrometer with thin carbon foils in the start detector
Detection efficiency of a heavy ion time‐of‐flight spectrometer with thin carbon foils in the start detector
复制标题
起始探测器中带有薄碳箔的重离子飞行时间谱仪的探测效率
DOI:
--
复制
发表时间:
1982
期刊:
影响因子:
--
通讯作者:
B. Wilken
中科院分区:
文献类型:
--
作者:
W. Stüdemann;B. Wilken
We report measurements of the effect of heavy‐ion scattering due to the passage through carbon foils with thicknesses of 4.5, 6.7, and 16.9 μg/cm2 in a wide energy range. For comparison of the results with theoretical predictions the actual foil thicknesses were determined by scaling energy loss measurements with tabulated stopping powers. Using these thicknesses the measurements in general agree well with theory. However, particularly for the 4.5 μg/cm2 foil, significant deviations from the specified thickness have been found very likely due to additional cover layers. The results are used to determine the loss in detection efficiency for low‐energy heavy ions in instruments based on the time‐of‐flight method with thin carbon foils acting as converters for secondary electrons.