Effects of cavitation-enhanced heating in high-intensity focused ultrasound treatment on shear wave imaging
Effects of cavitation-enhanced heating in high-intensity focused ultrasound treatment on shear wave imaging
复制标题
DOI:
10.7567/jjap.54.07hf11
复制
发表时间:
2015-07-01
影响因子:
1.5
通讯作者:
Umemura, Shin-ichiro
中科院分区:
文献类型:
--
作者:
Iwasaki, Ryosuke;Nagaoka, Ryo;Umemura, Shin-ichiro
High-intensity focused ultrasound (HIFU) therapy is a less invasive method of cancer treatment, in which ultrasound is generated outside the body and focused at the tumor tissue to be thermally coagulated. To enhance the safety, accuracy, and efficiency of HIFU therapy, "multiple-triggered HIFU" has been proposed as a method of cavitation-enhanced heating to shorten treatment time. In this study, we also propose shear wave elastography (SWE) to noninvasively monitor the cavitation-enhanced heating. Results show that the increase in shear wave velocity was observed in the coagulation area, but it was significantly slower when cavitation occurred. This suggests that the cavitation-enhanced heating requires a significantly longer cooling time before the accurate measurement of shear modulus than heating without generating bubbles. (C) 2015 The Japan Society of Applied Physics