CONTRALATERAL EFFECTS IN THE ELECTRO‐OCULOGRAM

CONTRALATERAL EFFECTS IN THE ELECTRO‐OCULOGRAM
复制标题

眼电图中的对侧效应

DOI:
10.1111/j.1755-3768.1974.tb01755.x
复制
发表时间:
1974
影响因子:
3.4
通讯作者:
A. Pinckers
A. Pinckers
中科院分区:
医学3区
文献类型:
--
作者:
J. Thijssen;A. Pinckers

文献摘要

被引文献

相似文献

从EOG数据的对侧效应的推导。数学程序可以适用于特殊的实验条件,如单侧咬合和不对称固定,或例如单侧眼球摘除的患者。实验结果显著地显示了眼睛的站立电位的对侧扩散,并且用因子f表示。该因子约为15%,代表对侧测量的同侧电位的分数。提出了一种在单侧损伤情况下对EOG数据进行校正的方法。
The derivation of the contralateral effect from EOG data is presented. The mathematical procedure can be adapted to special experimental conditions, such as unilateral occlusion, and asymmetric fixation, or for instance to a group of patients with unilateral enucleation. The contralateral spread of the standing potential of the eye is significantly shown by the experimental results, and is expressed by a factor f. This factor is of the order of 15% and represents the fraction of the ipsilateral potential that is measured contralaterally. A method is given to correct the EOG data in case of unilateral impairment.