Dosimetric equivalence of nonstandard HDR brachytherapy catheter patterns
Dosimetric equivalence of nonstandard HDR brachytherapy catheter patterns
复制标题
DOI:
10.1118/1.3041166
复制
发表时间:
2009-01-01
期刊:
影响因子:
3.8
通讯作者:
Pouliot, J.
中科院分区:
文献类型:
--
作者:
Cunha, J. A. M.;Hsu, I-C.;Pouliot, J.
Purpose: To determine whether alternative high dose rate prostate brachytherapy catheter patterns can result in similar or improved dose distributions while providing better access and reducing trauma. Materials and Methods: Standard prostate cancer high dose rate brachytherapy uses a regular grid of parallel needle positions to guide the catheter insertion. This geometry does not easily allow the physician to avoid piercing the critical structures near the penile bulb nor does it provide position flexibility in the case of pubic arch interference. This study used CT datasets with 3 mm slice spacing from ten previously treated patients and digitized new catheters following three hypothetical catheter patterns: conical, bi-conical, and fireworks. The conical patterns were used to accommodate a robotic delivery using a single entry point. The bi-conical and fireworks patterns were specifically designed to avoid the critical structures near the penile bulb. For each catheter distribution, a plan was optimized with the inverse planning algorithm, IPSA T, and compared with the plan used for treatment. Irrelevant of catheter geometry, a plan must fulfill the RTOG-0321 dose criteria for target dose coverage (V(100)(Prostate) > 90%) and organ-at-risk dose sparing (V(75)(Bladder) < 1 cc, V(75)(Rectum) < 1 cc, V(125)(Urethra)