Measurement of laser diode astigmatism
Measurement of laser diode astigmatism
复制标题
激光二极管像散的测量
DOI:
--
复制
发表时间:
1997
期刊:
影响因子:
--
通讯作者:
Haiyi Sun
中科院分区:
文献类型:
--
作者:
Haiyi Sun
Theoretical and experimental studies are presented on the two commonly used laser diode astigmatism measurement methods: the moving-diode method and the moving-profiler method. Both methods use lens(es) to focus the laser beam and a camera (or beam profiler) to locate the focal point of the focused beam. It is theoretically shown that the moving-diode method is valid when the distance between the diode and the lens is much larger than the lens focal length and the camera is positioned at a corresponding distance to the lens, and that virtually no calculation is required in a measurement. It is theoretically and experimentally shown that the moving-profiler method is valid when the measured data are elaborately processed using Gaussian optics and that the previous measurement with the measured data being processed by geometric optics was wrong.