MRI-Compatible Intensity-Modulated Force Sensor for Cardiac Catheterization Procedures
MRI-Compatible Intensity-Modulated Force Sensor for Cardiac Catheterization Procedures
复制标题
DOI:
10.1109/tbme.2010.2095853
复制
发表时间:
2011-03-01
影响因子:
4.6
通讯作者:
Althoefer, Kaspar
中科院分区:
文献类型:
--
作者:
Polygerinos, Panagiotis;Ataollahi, Asghar;Althoefer, Kaspar
This paper presents a novel, magnetic resonance imaging (MRI)-compatible, force sensor suitable for cardiac catheterization procedures. The miniature, fiber-optic sensor is integrated with the tip of a catheter to allow the detection of interaction forces with the cardiac walls. The optical fiber light intensity is modulated when a force acting at the catheter tip deforms an elastic element, which, in turn, varies the distance between a reflector and the optical fiber. The tip sensor has an external diameter of 9 Fr (3mm) and can be used during cardiac catheterization procedures. The sensor is able to measure forces in the range of 0-0.85 N, with relatively small hysteresis. A nonlinear method for calibration is used and real-time MRI in vivo experiments are carried out, to prove the feasibility of this low-cost sensor, enabling the detection of catheter-tip contact forces under dynamic conditions.