Single-step stereolithography of complex anatomical models for optical flow measurements

Single-step stereolithography of complex anatomical models for optical flow measurements
复制标题

DOI:
10.1115/1.1835367
复制
发表时间:
2005-02-01
影响因子:
1.7
通讯作者:
Yoganathan, AP
Yoganathan, AP
中科院分区:
工程技术4区
文献类型:
--
作者:
de Zélicourt, D;Pekkan, K;Yoganathan, AP

文献摘要

被引文献

相似文献

Transparent stereolithographic rapid prototyping (RP) technology has already demonstrated in literature to be a practical model construction tool for optical flow measurements such as digital particle image velocimetry (DPIV), laser doppler velocimetry (LDV), and flow visualization. Here, we employ recently available transparent RP resins and eliminate time-consuming casting and chemical curing steps from the traditional approach. This note details our methodology with relevant material properties and highlights its advantages. Stereolithographic model printing with our procedure is now a direct single-step process, enabling faster geometric replication of complex computational fluid dynamics (CFD) models for exact experimental validation studies. This methodology is specifically applied to the in vitro flow modeling of patient-specific total cavopulmonary connection (TCPC) morphologies. The effect of RP machining grooves, surface quality, and hydrodynamic performance measurements as compared with the smooth glass models are also quantified.