Design and performance of an imaging plate system for X-ray diffraction study

Design and performance of an imaging plate system for X-ray diffraction study
复制标题

DOI:
10.1016/0168-9002(88)90458-5
复制
发表时间:
1988-04
影响因子:
1.4
通讯作者:
Y. Amemiya;T. Matsushita;A. Nakagawa;Y. Satow;J. Miyahara;J. Chikawa
Y. Amemiya;T. Matsushita;A. Nakagawa;Y. Satow;J. Miyahara;J. Chikawa
中科院分区:
物理与天体物理3区
文献类型:
--
作者:
Y. Amemiya;T. Matsushita;A. Nakagawa;Y. Satow;J. Miyahara;J. Chikawa

文献摘要

被引文献

相似文献

通过改进鼓式扫描仪,构建了一种新的BaFBr: Eu2+光刺激荧光粉屏幕(成像板)读出系统,并对其设计进行了优化,以适应x射线衍射和散射应用。利用12位a /D转换器实现了低于20 keV的高探测量子效率、小像素尺寸(25 μm × 25 μm)、低量化噪声(0.22%),并通过使用两个光电倍增管实现了覆盖固有动态范围(1:10 05)的光激发发光能力。该系统被用于几个同步辐射实验:蛋白质晶体的劳埃衍射,单个肌肉纤维的小角度衍射,金刚石铁质细胞中晶体的粉末衍射,以及在拉伸过程中合成聚合物的时间分辨小角度x射线散射。
A new readout system for a BaFBr: Eu2+photostimulable phosphor screen (imaging plate) was constructed by modifying a drum scanner, with a design optimized for X-ray diffraction and scattering applications. An effort was made to achieve a high detective quantum efficiency below 20 keV, a small pixel size (25 μm × 25 μm), a low quantization noise (0.22%) using 12-bit A/D converters, and the capability to cover an inherent dynamic range (1:105) of the photostimulated luminescence by using two photomultiplier tubes. This system is being used in several synchrotron radiation experiments: Laue diffraction of protein crystals, small angle diffraction from a single muscle fiber, powder diffraction from crystals in a diamond anvil cell, and time-resolved small-angle X-ray scattering from a synthetic polymer during stretching.