Design and performance of an imaging plate system for X-ray diffraction study
Design and performance of an imaging plate system for X-ray diffraction study
复制标题
DOI:
10.1016/0168-9002(88)90458-5
复制
发表时间:
1988-04
影响因子:
1.4
通讯作者:
Y. Amemiya;T. Matsushita;A. Nakagawa;Y. Satow;J. Miyahara;J. Chikawa
中科院分区:
文献类型:
--
作者:
Y. Amemiya;T. Matsushita;A. Nakagawa;Y. Satow;J. Miyahara;J. Chikawa
A new readout system for a BaFBr: Eu2+photostimulable phosphor screen (imaging plate) was constructed by modifying a drum scanner, with a design optimized for X-ray diffraction and scattering applications. An effort was made to achieve a high detective quantum efficiency below 20 keV, a small pixel size (25 μm × 25 μm), a low quantization noise (0.22%) using 12-bit A/D converters, and the capability to cover an inherent dynamic range (1:105) of the photostimulated luminescence by using two photomultiplier tubes. This system is being used in several synchrotron radiation experiments: Laue diffraction of protein crystals, small angle diffraction from a single muscle fiber, powder diffraction from crystals in a diamond anvil cell, and time-resolved small-angle X-ray scattering from a synthetic polymer during stretching.