Shielded open-circuited sample holder for dielectric measurements of solids and liquids
Shielded open-circuited sample holder for dielectric measurements of solids and liquids
复制标题
用于固体和液体介电测量的屏蔽开路样品架
DOI:
10.1109/19.744172
复制
发表时间:
1998
期刊:
影响因子:
--
通讯作者:
C. Jones
中科院分区:
文献类型:
--
作者:
J. Baker;M. Janezic;C. Jones
This paper presents a technique for permittivity measurements that can be used for either liquids or solid coaxial samples from near dc to gigahertz frequencies. The method uses both a full-mode model of a coaxial open-circuited termination and a capacitive method for permittivity determination. The method requires the use of an impedance bridge and a network analyzer. Measurements indicate good accuracy for the permittivity over a very wide band of frequencies. For high permittivity materials the sample surfaces must be metallized. An uncertainty analysis is also presented.