Asymmetric Interdigitated Electrodes for Amperometric Detection of Soluble Products
Asymmetric Interdigitated Electrodes for Amperometric Detection of Soluble Products
复制标题
DOI:
10.1149/1945-7111/ac001c
复制
发表时间:
2021-05
影响因子:
3.9
通讯作者:
Sophia E. Lee;Maureen H. Tang
中科院分区:
文献类型:
--
作者:
Sophia E. Lee;Maureen H. Tang
A new electroanalytical design based on asymmetric interdigitated arrays (IDA) is presented and the effects of asymmetry on the device performance are characterized electrochemically. Varying the collector and generator band widths independently of each other tunes the collection efficiency and redox cycling-induced feedback. These arrays are able to provide low feedback (< 10%) while maintaining moderate collection efficiency (25%–40%). Behavior is evaluated experimentally and using a numerical model. The application of the device to detect soluble electrolyte degradation products in nonaqueous lithium-ion and sodium-ion battery electrolytes is demonstrated.