Artifacts Related to Sample Introduction in Capillary Gel Electrophoresis Affecting Separation Performance and Quantitation
Artifacts Related to Sample Introduction in Capillary Gel Electrophoresis Affecting Separation Performance and Quantitation
复制标题
与影响分离性能和定量的毛细管凝胶电泳进样相关的人为因素
DOI:
10.1021/ac00109a056
复制
发表时间:
1995
影响因子:
7.4
通讯作者:
H. Schwartz
中科院分区:
文献类型:
--
作者:
A. Guttman;H. Schwartz
Two injection-related artifacts with capillary gel electrophoresis (CGE) are reported. The first occurs with consecutive injections from the same, low-volume (10-200 μL), aqueous sample : progressively smaller peak heights are obtained with each injection. This phenomenon is explained in qualitative terms by examining the electrochemical processes occurring during the electrokinetic sample introduction. A simple solution to this problem is to perform an intermediate electrokinetic injection from a water vial prior to the sample injection. This two-step injection procedure results not only in dramatically increased precision (important for quantitative studies) but also in increased sample loading. The second artifact observed in CGE is related to the physical shape of the inlet of the gel-filled capillary : employing an oblique edge capillary results in poor performance, peak distortion, and loss in resolution.