A Novel Approach to Contamination Suppression in Transmission Detectors for Radiotherapy
A Novel Approach to Contamination Suppression in Transmission Detectors for Radiotherapy
复制标题
DOI:
10.1109/trpms.2020.2995059
复制
发表时间:
2020-09-01
影响因子:
4.4
通讯作者:
Pritchard, J.
中科院分区:
文献类型:
--
作者:
Beck, L.;Velthuis, J. J.;Pritchard, J.
The current trend in X-ray radiotherapy is to treat cancers that are in difficult locations in the body using beams with a complex intensity profile. Intensity-modulated radiotherapy (IMRT) is a treatment which improves the dose distribution to the tumor whilst reducing the dose to healthy tissue. Such treatments administer a larger dose per treatment fraction and hence require more complex methods to verify the accuracy of the treatment delivery. Measuring beam intensity fluctuations is difficult as the beam is heavily distorted after leaving the patient and transmission detectors will attenuate the beam and change the energy spectrum of the beam. Monolithic active pixel sensors (MAPSs) are ideal solid-state detectors to measure the 2-D beam profile of a radiotherapy beam upstream of the patient. MAPS sensors can be made very thin (similar to 30 mu m) with still very good signal-to-noise performance. This means that the beam would pass through the sensor virtually undisturbed (