Automated suprathreshold static perimetry.

Automated suprathreshold static perimetry.
复制标题

自动阈上静态视野检查。

DOI:
10.1016/0002-9394(80)90296-2
复制
发表时间:
1980
影响因子:
4.2
通讯作者:
J. Keltner
J. Keltner
中科院分区:
医学1区
文献类型:
--
作者:
Chris A. Johnson;J. Keltner

文献摘要

被引文献

相似文献

对1,019只患有各种视力障碍的患者进行了自动视野检查(Fieldmaster)和手动动态视野检查。在适当的测试条件下,自动Fieldmaster评估显示,视野缺陷检测的检测率为96.1%,误警率为4.7%。对168只眼的初步结果显示,Fieldmaster的连续随访测试正确地识别了超过94%(36只眼中的34只眼)视野改变的视野丧失的进展或消退。Fieldmaster正确地确定了99%(132只眼中的131只)视野稳定的眼睛的视野特征没有改变。这些数据表明,Fieldmaster是一种检测视野缺陷的优秀设备,也具有监测视野丧失进展或回归的能力。
Automated (Fieldmaster) and manual kinetic perimetry were performed in 1,019 eyes with various visual disorders. tunder appropriate test conditions, automated Fieldmaster evaluations showed a 96.1% detection rate and a 4.7% false alarm rate for detection of visual field defects. Preliminary results for 168 eyes showed that sequential follow-up testing with the Fieldmaster correctly identified progression or regression of visual field loss in more than 94%(34 of 36) of the eyes with visual field changes. The Fieldmaster correctly determined no change in visual field characteristics for 99%(131 of 132) of the eyes with stable visual fields. These data indicate that the Fieldmaster is an excellent device for detecting visual field defects, and also has some capabilities for monitoring progression or regression of visual field loss.