Nuclear morphology measurements using Fourier domain low coherence interferometry
Nuclear morphology measurements using Fourier domain low coherence interferometry
复制标题
DOI:
10.1364/opex.13.004693
复制
发表时间:
2005-06-13
期刊:
影响因子:
3.8
通讯作者:
Wax, A
中科院分区:
文献类型:
--
作者:
Graf, RN;Wax, A
We present a new common path configuration Fourier domain low coherence interferometry (fLCI) optical system and demonstrate its capabilities by presenting results which determine the size of cell nuclei in a monolayer of T84 epithelial cells. The optical system uses a white light source in a modified Michelson interferometer and a spectrograph for detection of the mixed signal and reference fields. Depth resolution is obtained from the Fourier transform of the measured spectrum which provides the axial spatial cross-correlation between the signal and reference fields. The spectral dependence of scattering by the samples is determined by windowing the spectrum to measure the scattering amplitude as a function of wavenumber. We present evidence that fLCI accurately measures the longitudinal profile of cell nuclei rather than the transverse profile. (C) 2005 Optical Society of America.