CINERADIOGRAPHIC FRAME RATE SELECTION FOR LEFT VENTRICULAR VOLUMETRY
CINERADIOGRAPHIC FRAME RATE SELECTION FOR LEFT VENTRICULAR VOLUMETRY
复制标题
DOI:
10.1148/96.3.587
复制
发表时间:
1970-01-01
期刊:
影响因子:
19.7
通讯作者:
LYNCH, PR
中科院分区:
文献类型:
--
作者:
FREEMAN, E;ZISKIN, MC;LYNCH, PR
The left ventricular volume curve was plotted from biplane cineangiograms at 270 frames per second, using Simpson's rule to measure the volume. Fourier transformation was applied to obtain the frequency spectrum of the curve. Volume curves at slower rates were produced. Curves with less than 5% error can be generated at 15 frames per second because of the low frequency content. This rate of filming was adequate to measure left ventricular volume by biplane cineradiography.