Simultaneous fat saturation and magnetization transfer contrast imaging with steady-state incoherent sequences.
Simultaneous fat saturation and magnetization transfer contrast imaging with steady-state incoherent sequences.
复制标题
使用稳态不相干序列同时进行脂肪饱和度和磁化传递对比成像。
DOI:
10.1002/mrm.25475
复制
发表时间:
2015
影响因子:
3.3
通讯作者:
Noll,DouglasC
中科院分区:
文献类型:
--
作者:
Zhao,Feng;Nielsen,Jon-Fredrik;Swanson,ScottD;Fessler,JeffreyA;Noll,DouglasC
PurposeThis work combines ann‐dimensional fat sat(uration) radiofrequency (RF) pulse with steady‐state incoherent (SSI) pulse sequences, e.g., spoiled gradient‐echo sequence, to simultaneously produceB0insensitive fat suppression and magnetization transfer (MT) contrast. This pulse is then referred to as “fat sat and MT contrast pulse.”TheoryWe discuss the features of the fat sat and MT contrast pulse and the MT sensitivities of the SSI sequences when combining with fat sat. Moreover, we also introduce an adapted RF spoiling scheme for SSI sequences with fat sat. Methods: Simulations and phantom experiments were conducted to demonstrate the adapted RF spoiling. Fat suppression and MT effects are shown in 3T phantom experiments and in vivo experiments, including brain imaging, cartilage imaging, and angiography.ResultsTo ensure that the sequence reaches steady state, the adapted RF spoiling is required for fat sat SSI sequences. Fat sat and MT contrast pulse works robustly with field inhomogeneity and also produces MT contrasts.ConclusionSSI sequences with fat sat and MT contrast pulse and adapted RF spoiling can robustly produce fat suppressed and MT contrast images in the presence of field inhomogeneity. Magn Reson Med 74:739–746, 2015. © 2014 Wiley Periodicals, Inc.