Cross-talk in otoacoustic emission probes.
Cross-talk in otoacoustic emission probes.
复制标题
耳声发射探头中的串扰。
DOI:
10.1097/00003446-199504000-00002
复制
发表时间:
1995
期刊:
影响因子:
3.7
通讯作者:
Siegel,JH
中科院分区:
文献类型:
--
作者:
Siegel,JH
ObjectiveSeveral commercial otoacoustic emission probes (Virtual 330, Etymbtic Research ER-10 and ER-IOB, and Otodynamics IL088 and IL092) were tested for internal coupling (cross-talk) between the sound source and probe microphone.DesignThe response of the probe under test was measured, in a cavity that simulates the human ear canal, first in its normal configuration and then with either the sound source tube or the microphone inlet blocked with modelling clay. Cross-talk was indicated if the pressure failed to drop substantially in the “blocked” condition.ResultsCross-talk in the Virtual 330 and Etymutic ER-IOB was found to be severe enough to compromise the accuracy of “in the ear” acoustic calibrations. In both cases the cause was the radiation of sound through the walls of the flexible silicone tubing used to conduct the stimuli through the body of the probe. Substituting harder-walled sound source tubing reduced cross-talk to levels acceptable for routine application. The probes used in the Otodynamics IL088 and KO92 systems and the original Etymbtic ERIO probes all have acceptably low cross-talk as supplied from the manufacturers.ConclusionsFor the two probes where excessive cross-talk was found, most of the cross-talk was acoustic and or mechanical, rather than electromagnetic. Artifacts due to cross-talk have undoubtedly contaminated data collected with these units. Manufacturers of emission probes are urged to adopt a standard for specifying cross-talk levels and include this specification in product literature.