STOPPING POWER DATA FOR HIGH-ENERGY PHOTON BEAMS
STOPPING POWER DATA FOR HIGH-ENERGY PHOTON BEAMS
复制标题
DOI:
10.1088/0031-9155/31/8/002
复制
发表时间:
1986-08-01
影响因子:
3.5
通讯作者:
BRAHME, A
中科院分区:
文献类型:
--
作者:
ANDREO, P;BRAHME, A
Spencer-Attix stopping power ratios for the dosimetry of high-energy photon beams used in radiation therapy have been calculated using the Monte Carlo method. The stopping power ratios are calculated in a more consistent way than previously and are given a function of the attenuation properties of the beam. The dependence of the stopping power ratio on the electron contamination of the beam as well as on depth and field size has also been investigated. Results are compared with stopping power ratios recommended in different dosimetry protocols and to experimental results. The agreement with most dosimetry protocols is within about one per cent with recent experimental data is better than half a per cent.