On-line monitoring of the methanol concentration in Pichia pastoris cultures producing an heterologous lipase by sequential injection analysis
On-line monitoring of the methanol concentration in Pichia pastoris cultures producing an heterologous lipase by sequential injection analysis
复制标题
DOI:
10.1023/a:1026228518418
复制
发表时间:
2003-11
影响因子:
2.7
通讯作者:
A. Surribas;O. Cós;J. Montesinos;F. Valero
中科院分区:
文献类型:
--
作者:
A. Surribas;O. Cós;J. Montesinos;F. Valero
An automated sequential injection analysis (SIA) system using stop-flow technique was developed to determine methanol concentration by means of the enzymatic reactions of alcohol oxidase and peroxidase. Its application as an on-line device for monitoringPichia pastorisfermentations producing an heterologous protein was demonstrated. Linear response, observed up to 2 g l−1, was reached by including a dilution chamber in the SIA manifold. The sampling frequency was 7 analyses per hour with a relative standard deviation lower than 4%.