How few cancer cells can be detected by positron emission tomography? A frequent question addressed by an in vitro study
How few cancer cells can be detected by positron emission tomography? A frequent question addressed by an in vitro study
复制标题
DOI:
10.1007/s00259-005-0038-6
复制
发表时间:
2006-06-01
影响因子:
9.1
通讯作者:
Kristjansen, PEG
中科院分区:
文献类型:
--
作者:
Fischer, BM;Olsen, MWB;Kristjansen, PEG
Purpose: Positron emission tomography ( PET) has gained widespread use in cancer diagnosis and treatment, but how many malignant cells are required for a tumour to be detected by PET?Methods: Three human cancer cell lines [ glioblastoma and two subtypes of small cell lung cancer (SCLC)] in concentrations from 10(4) to 10(7) were seeded on six-well plates or plastic tubes and treated with [F-18] fluorodeoxyglucose (FDG) in vitro. FDG retention was measured in a PET/CT scanner and in a calibrated well counter. The clinical situation was simulated using a cylinder phantom with a background concentration of FDG.Results: The theoretical detection limit was found to be around 10(5) malignant cells. In a cylinder phantom the detection limit was increased by a factor of 10. The FDG retention by the glioblastoma cell line was significantly higher than the activity of the SCLC cell line. FDG retention measured by PET and a gamma counter was closely correlated to the number of cells and a linear relationship was found.Discussion: The detection limit of PET is in the magnitude of 10(5) to 10(6) malignant cells. The experimental set-up was robust and well suited as a platform for further investigations of factors influencing the detection limit of PET.