The delivery of intensity modulated radiotherapy to the breast using multiple static fields
The delivery of intensity modulated radiotherapy to the breast using multiple static fields
复制标题
DOI:
10.1016/s0167-8140(00)00263-2
复制
发表时间:
2000-10-01
影响因子:
5.7
通讯作者:
Yarnold, JR
中科院分区:
文献类型:
--
作者:
Evans, PM;Donovan, EM;Yarnold, JR
Background and purpose: To develop a method of using a multileaf collimator (MLC) to deliver intensity modulated radiotherapy (IMRT) for tangential breast fields, using an MLC to deliver a set of multiple static fields (MSFs).Materials and methods: An electronic portal imaging device (EPID) is used to obtain thickness maps of medial and lateral tangential breast fields. From these IMRT deliveries are designed to minimize the volume of breast above 105% of prescribed dose. The deliveries are universally-wedged beams augmented with a set of low dose shaped irradiations. Dosimetric and planning QA of this method has been compared with the standard, wedged treatment and the corresponding treatment using physical compensators. Several options for delivering the MSF treatment are presented.Results: The MSF technique was found to be superior to the standard technique (P value = 0.002) and comparable with the compensated technique. Both IMRT methods reduced the volume of breast above 105% dose from a mean value of 12.0% of the total breast volume to similar to2.8% of the total breast volume.Conclusions: This MSF method may be used to reduce the high dose volume in tangential breast irradiation significantly. This may have consequences for long-term side effects, particularly cosmesis. (C) 2000 Elsevier Science Ireland Ltd. All rights reserved.