Low-resolution colonic manometry leads to a gross misinterpretation of the frequency and polarity of propagating sequences: Initial results from fiber-optic high-resolution manometry studies
Low-resolution colonic manometry leads to a gross misinterpretation of the frequency and polarity of propagating sequences: Initial results from fiber-optic high-resolution manometry studies
复制标题
DOI:
10.1111/nmo.12170
复制
发表时间:
2013-10-01
影响因子:
3.5
通讯作者:
Arkwright, J. W.
中科院分区:
文献类型:
--
作者:
Dinning, P. G.;Wiklendt, L.;Arkwright, J. W.
BackgroundHigh-resolution manometry catheters are now being used to record colonic motility. The aim of this study was to determine the influence of pressure sensor spacing on our ability to identify colonic propagating sequences (PS).MethodsFiber-optic catheters containing 72-90 sensors spaced at 1cm intervals were placed colonoscopically to the cecum in 11 patients with proven slow transit constipation, 11 patients with neurogenic fecal incontinence and nine healthy subjects. A 2h section of trace from each subject was analyzed. Using the 1cm spaced data as the gold standard, each data set was then sub-sampled, by dropping channels from the data set to simulate sensor spacing of 10, 7, 5, 3, and 2cm. In blinded fashion, antegrade and retrograde PS were quantified at each test sensor spacing. The data were compared to the PSs identified in the corresponding gold standard data set.Key ResultsIn all subject groups as sensor spacing increased; (i) the frequency of identified antegrade and retrograde PSs decreased (P