On the extinction coefficient in confocal polarization microscopy
On the extinction coefficient in confocal polarization microscopy
复制标题
共焦偏振显微镜中的消光系数
DOI:
10.1111/j.1365-2818.1995.tb03636.x
复制
发表时间:
1995
影响因子:
2
通讯作者:
R. Juškaitis
中科院分区:
文献类型:
--
作者:
T. Wilson;R. Juškaitis
The extinction coefficient in conventional polarized light microscopes is non‐zero, even with perfect polars, solely because of the image formation process in these instruments. The imaging in confocal microscopes is different from conventional instruments and it is shown that in this case an infinite extinction coefficient results in the ideal case whereas a finite value is always to be expected with conventional instruments. Images of the same microelectronic specimen taken in both conventional and confocal polarized light microscopes are compared.