Measuring Apparent Trabecular Structure With pQCT: A Comparison With HR-pQCT
Measuring Apparent Trabecular Structure With pQCT: A Comparison With HR-pQCT
复制标题
DOI:
10.1016/j.jocd.2013.03.002
复制
发表时间:
2014-01-01
影响因子:
2.5
通讯作者:
Giangregorio, Lora
中科院分区:
文献类型:
--
作者:
Lala, Deena;Cheung, Angela M.;Giangregorio, Lora
We evaluated how comparable peripheral quantitative computed tomography (pQCT) measurements of cortical thickness, density, and apparent trabecular structure at the ultradistal tibia were with those measured with high-resolution pQCT (HR-pQCT). We also examined whether the accuracy of the pQCT-based trabecular and cortical measurements improved with reductions in slice thickness from the standard 2.2 mm to 1.1 and 0.6 mm. We immersed 15 dry tibia specimens in saline in a sealed cylinder and scanned 22 5 mm from the distal tibia plateau using pQCT and HR-pQCT. pQCT underestimated cortical thickness by Stratec (CThStratec) and trabecular spacing (Tb.Sp) by 21.4% and 72.9%, whereas bone volume to total volume (BV/TV) and cortical density (CDen) were overestimated by 265.8% and 13.1%, respectively. Measurements of trabecular volumetric bone mineral density, trabecular area, total area, cortical thickness by custom software were comparable, but for CThStratec, Tb.Sp, BV/TV, and CDen, the differences between imaging devices varied with magnitude of the estimate. We recommend that researchers or clinicians interested in using pQCT to measure apparent trabecular structure or cortical thickness at the epiphyses, or in comparing findings from different devices, be aware of the differences between HR-pQCT and pQCT.