Dose and image quality of electron-beam CT compared with spiral CT.
Dose and image quality of electron-beam CT compared with spiral CT.
复制标题
电子束CT与螺旋CT的剂量和图像质量比较。
作者:
Gerald Weisser;Lehmann Kj;Roland J. Scheck;Eva Coppenrath;Max Georgi
RATIONALE AND OBJECTIVES
To examine dose and image quality of electron-beam CT (EBCT) with continuous volume scan versus spiral CT.
METHODS
An EBCT scanner was compared with a spiral CT (SCT) scanner. Three phantoms were used to measure low-contrast resolution, high-contrast resolution, slice width, and dose.
RESULTS
The EBCT scans showed 30% lower high-contrast resolution for most settings. The dose was comparable to that of spiral CT with 3 mm collimation and 76%/106% higher with EBCT for 1.5 mm/6 mm collimation. Low-contrast resolution was comparable to that of spiral CT using 3 mm collimation, slightly worse for 1.5 mm, and bad for 6 mm EBCT collimation (four times higher dose to reach comparable contrast-to-noise ratio).
CONCLUSIONS
Significant restrictions were found using EBCT with continuous volume scan. The authors found that 3 mm collimation can yield acceptable high-contrast resolution and good low-contrast resolution compared with spiral CT. The use of 6 mm or 1.5 mm collimation needs to be restricted to selected cases.