A HIGH PRESSURE XENON SELF-TRIGGERED SCINTILLATION DRIFT CHAMBER WITH 3D SENSITIVITY IN THE RANGE OF 20-140 KEV DEPOSITED ENERGY
A HIGH PRESSURE XENON SELF-TRIGGERED SCINTILLATION DRIFT CHAMBER WITH 3D SENSITIVITY IN THE RANGE OF 20-140 KEV DEPOSITED ENERGY
复制标题
3D灵敏度20-140 KEV沉积能量的高压氙自触发闪烁漂移室
DOI:
10.1016/s0168-9002(96)01035-2
复制
发表时间:
1997
影响因子:
1.4
通讯作者:
V. Morgunov
中科院分区:
文献类型:
--
作者:
A. Bolozdynya;V. Egorov;A. Koutchenkov;G. Safronov;G. Smirnov;S. Medved';V. Morgunov
A high pressure gas self-triggered scintillation drift chamber has been built with a working volume of 30 cm diameter and 4 cm depth, viewed through glass windows by an array of 19 PMTs. The detector tested at 9 bar xenon filling showed 2.7% FWHM energy resolution at 122 keV over a 22 cm field of view and three-dimensional sensitivity with X,Y resolution of 3.7 mm FWHM and Z resolution of 0.6 mm FWHM for 60 keV photoabsorbed γ-rays. The readout system of the detector is able to select multi-vertex events occurring in the sensitive volume due to γ-ray absorption and discriminate fluorescent photons, Compton scattering and photoabsorption vertices. The energy resolution in “fluorescent photon gated mode” has been measured as 2.2% FWHM at 122 keV γ-ray. Some possible applications of the detector are discussed.