Size-of-source Effect in Infrared Thermometers with Direct Reading of Temperature
Size-of-source Effect in Infrared Thermometers with Direct Reading of Temperature
复制标题
DOI:
10.1007/s10765-017-2237-3
复制
发表时间:
2017-07-01
影响因子:
2.2
通讯作者:
Saunders, P.
中科院分区:
文献类型:
--
作者:
Manoi, A.;Saunders, P.
The size-of-source effect (SSE) for six infrared (IR) thermometers with direct reading of temperature was measured in this work. The alternative direct method for SSE determination, where the aperture size is fixed and the measurement distance is varied, was used in this study. The experimental equivalence between the usual and the alternative direct methods is presented. The magnitudes of the SSE for different types of IR thermometers were investigated. The maxima of the SSE were found to be up to 5 %, 8 %, and 28 % for focusable, closed-focus, and open-focus thermometers, respectively. At 275 degrees C, an SSE of 28 % corresponds to 52 degrees C, indicating the severe effect on the accuracy of this type of IR thermometer. A method to realize the calibration conditions used by the manufacturer, in terms of aperture size and measurement distance, is discussed and validated by experimental results. This study would be of benefit to users in choosing the best IR thermometer to match their work and for calibration laboratories in selecting the technique most suitable for determining the SSE.