A low-cost miniaturized potentiostat for point-of-care diagnosis
A low-cost miniaturized potentiostat for point-of-care diagnosis
复制标题
DOI:
10.1016/j.bios.2014.06.053
复制
发表时间:
2014-12-15
影响因子:
12.6
通讯作者:
Bhansali, Shekhar
中科院分区:
文献类型:
--
作者:
Cruz, Andres Felipe Diaz;Norena, Nicolas;Bhansali, Shekhar
This paper presents a novel approach of using a miniaturized potentiostat (M-P) chip (LMP91000) to perform full range cyclic voltammetry (CV) measurements for the detection of biomarkers. The LMP91000 evaluation board was reconfigured to perform three-electrode CV measurements in order to achieve electrochemical cortisol immunosensing. The microelectrodes for cortisol estimation were fabricated by immobilizing monoclonal anti-cortisol antibody (Anti-M-Cab) onto self-assembled monolayer (SAM) modified Au microelectrodes. The results obtained using the M-P were compared to those obtained using a conventional potentiostat. The M-P was successful in measuring cortisol levels in the range of pM. The outcomes of the studies suggest that M-P can effectively perform biochemical measurements on three electrode systems, enabling the development of miniature systems for point-of-care (POC) diagnosis. (C) 2014 Published by Elsevier B.V.