Comprehensive Evaluation of Compendial USP, BacT/Alert Dual-T, and Bactec FX for Detection of Product Sterility Testing Contaminants
Comprehensive Evaluation of Compendial USP, BacT/Alert Dual-T, and Bactec FX for Detection of Product Sterility Testing Contaminants
复制标题
DOI:
10.1128/jcm.01548-18
复制
发表时间:
2019-02-01
影响因子:
9.4
通讯作者:
Lau, Anna F.
中科院分区:
文献类型:
--
作者:
England, Matthew R.;Stock, Frida;Lau, Anna F.
The emergence of cell therapy programs in large academic centers has led to an increasing demand for clinical laboratories to assist with product sterility testing. Automated blood culture systems have shown promise as alternatives to the manual USP compendial method, but current published data are limited by small organism test sets, particularly for molds. In 2015, failure of the Bactec FX system to detect mold contamination in two products prompted us to evaluate three test systems (compendial USP, Bactec FX, and BacT/Alert Dual-T) over seven different culture combinations, using 118 challenge organisms representative of the NIH current good manufacturing practice (cGMP) environment. At