In silico study of low-frequency transcranial ultrasound fields in acute ischemic stroke patients.

In silico study of low-frequency transcranial ultrasound fields in acute ischemic stroke patients.
复制标题

DOI:
10.1016/j.ultrasmedbio.2013.12.025
复制
发表时间:
2014-06
影响因子:
2.9
通讯作者:
Holland, Christy K.
Holland, Christy K.
中科院分区:
医学3区
文献类型:
--
作者:
Bouchoux, Guillaume;Shivashankar, Ravishankar;Abruzzo, Todd A.;Holland, Christy K.

文献摘要

参考文献

被引文献

相似文献

亚兆赫范围内的超声波可增强溶栓作用,可用于缺血性中风患者的经颅治疗。需要评估经颅电击的一致性。对20名患者进行了基于CT扫描的声学和热学模拟。无聚焦的120 kHz换能器允许均匀地电击血栓,并根据外部地标进行定位,类似于基于CT数据的优化放置。在500 kHz弱聚焦传感器的情况下,基于里程碑的定位表现不佳。预测的患者间原位声压变化与优化放置的两种探头相似(18.0-26.4%相对标准偏差)。120 kHz和500 kHz换能器模拟的介入组织的最大声压分别是血栓内压力的2.6±0.6和2.0±0.7倍。1W/cm2辐照血栓可使120 kHz换能器骨温度升高3.8±2.2℃,500 kHz换能器骨温度升高13.4±3.3℃。对于120 kHz换能器,预计对侧局部最大值可达目标区压力幅度的1.1倍。我们建立了两种传感器放置方法,一种是基于头部CT的分析,另一种是使用简单的外部可见地标。这两种方法都可以使血栓始终如一地精神化。
Ultrasound in the sub-megahertz range enhances thrombolysis and may be applied transcranially to ischemic stroke patients. The consistency of transcranial insonification needs to be evaluated. Acoustic and thermal simulations based on computed-tomography (CT) scans of 20 patients were performed. An unfocused 120-kHz transducer allowed homogeneous insonification of the thrombus, and positioning based on external landmarks performed similarly to an optimized placement based on CT data. With a weakly focused 500-kHz transducer, the landmark-based positioning underperformed. The predicted inter-patient variation of in situ acoustic pressure was similar with both transducers for the optimized placement (18.0–26.4% relative standard deviation). The simulated maximum acoustic pressure in intervening tissues was 2.6±0.6 and 2.0±0.7 times the pressure in the thrombus for the 120-kHz and 500-kHz transducers, respectively. A 1 W/cm2 insonification of the thrombus caused a 3.8±2.2°C temperature increase in the bone for the 120-kHz transducer, and a 13.4±3.3°C increase for the 500-kHz transducer. Contralateral local maxima up to 1.1 times the pressure amplitude in the targeted zone were predicted for the 120-kHz transducer. We established two transducer placement approaches, one based on analysis of a head CT and the other using simple external, visible landmarks. Both approaches allowed consistent insonification of the thrombus.
DOI: 10.1121/1.381852
发表时间: 1978-01-01
影响因子: 2.4
作者:
FRY, FJ;BARGER, JE
通讯作者: BARGER, JE
DOI: 10.1088/0031-9155/54/9/001
发表时间: 2009-05-07
影响因子: 3.5
作者:
Marquet, F.;Pernot, M.;Fink, M.
通讯作者: Fink, M.
DOI: 10.1158/0008-5472.can-12-0128
发表时间: 2012-07-15
期刊: Cancer research
影响因子: 11.2
作者:
McDannold N;Arvanitis CD;Vykhodtseva N;Livingstone MS
通讯作者: Livingstone MS
DOI: 10.1161/str.0b013e318284056a
发表时间: 2013-03-01
期刊: STROKE
影响因子: 8.3
作者:
Jauch, Edward C.;Saver, Jeffrey L.;Yonas, Howard
通讯作者: Yonas, Howard
DOI: 10.1002/mrm.20060
发表时间: 2004-05-01
影响因子: 3.3
作者:
McDannold, N;Vykhodtseva, N;Hynynen, K
通讯作者: Hynynen, K