Tracking errors in a prototype real-time tumour tracking system
Tracking errors in a prototype real-time tumour tracking system
复制标题
DOI:
10.1088/0031-9155/49/23/011
复制
发表时间:
2004-12-07
影响因子:
3.5
通讯作者:
Shirato, H
中科院分区:
文献类型:
--
作者:
Sharp, GC;Jiang, SB;Shirato, H
In motion-compensated radiation therapy, radio-opaque markers can be implanted in or near a tumour and tracked in real-time using fluoroscopic imaging. Tracking these implanted markers gives highly accurate position information, except when tracking fails due to poor or ambiguous imaging conditions. This study investigates methods for automatic detection of tracking errors, and assesses the frequency and impact of tracking errors on treatments using the prototype real-time tumour tracking system. We investigated four indicators for automatic detection of tracking errors. and found that the, distance between corresponding rays was most effective. We also found that tracking errors cause a loss of gating efficiency of between 7.6 and 10.2%. The incidence of treatment beam delivery during tracking errors was estimated at between 0.8% and 1.25%.