Microfluidic System for Automated Cell-based Assays.
Microfluidic System for Automated Cell-based Assays.
复制标题
DOI:
10.1016/j.jala.2007.07.001
复制
发表时间:
2007-12-01
期刊:
影响因子:
--
通讯作者:
Hung, Paul J
中科院分区:
文献类型:
--
作者:
Lee, Philip J;Ghorashian, Navid;Hung, Paul J
Microfluidic cell culture is a promising technology for applications in the drug screening industry. Key benefits include improved biological function, higher quality cell-based data, reduced reagent consumption, and lower cost. In this work, we demonstrate how a microfluidic cell culture design was adapted to be compatible with the standard 96-well plate format. Key design features include the elimination of tubing and connectors, the ability to maintain long term continuous perfusion cell culture using a passive gravity driven pump, and direct analysis on the outlet wells of the microfluidic plate. A single microfluidic culture plate contained 8 independent flow units, each with 10(4) cells at a flow rate of 50 mul/day (6 minute residence time). The cytotoxicity of the anti-cancer drug etoposide was measured on HeLa cells cultured in this format, using a commercial lactate dehydrogenase (LDH) plate reader assay. The integration of microfluidic cell culture methods with commercial automation capabilities offers an exciting opportunity for improved cell-based screening.