Improved MRI R2* Relaxometry of Iron-Loaded Liver with Noise Correction
Improved MRI R2* Relaxometry of Iron-Loaded Liver with Noise Correction
复制标题
DOI:
10.1002/mrm.24607
复制
发表时间:
2013-12-01
影响因子:
3.3
通讯作者:
Firmin, David N.
中科院分区:
文献类型:
--
作者:
Feng, Yanqiu;He, Taigang;Firmin, David N.
Accurate and reproducible MRI R-2* relaxometry for tissue iron quantification is important in managing transfusion-dependent patients. MRI data are often acquired using array coils and reconstructed by the root-sum-square algorithm, and as such, measured signals follow the noncentral chi distribution. In this study, two noise-corrected models were proposed for the liver R-2* quantification: fitting the signal to the first moment and fitting the squared signal to the second moment in the presence of the noncentral chi noise. These two models were compared with the widely implemented offset and truncation models on both simulation and in vivo data. The results demonstrated that the slow decay component of the liver R-2* was mainly caused by the noise. The offset model considerably overestimated R-2* values by incorrectly adding a constant to account for the slow decay component. The truncation model generally produced accurate R-2* measurements by only fitting the initial data well above the noise level to remove the major source of errors, but underestimated very high R-2* values due to the sequence limit of obtaining very short echo time images. Both the first and second-moment noise-corrected models constantly produced accurate and precise R-2* measurements by correctly addressing the noise problem. Magn Reson Med 70:1765-1774, 2013. (c) 2013 Wiley Periodicals, Inc.