The range of positrons in tissue is an important limitation to the ultimate spatial resolution achievable in positron emission tomography. In this work, a Fourier deconvolution technique is applied to remove range blurring in images taken by the Donner 600-crystal positron tomograph. Using phantom data, the authors found significant improvement in the image quality and the FWHM for both /sup 68/Ga and /sup 82/Rb. These were successfully corrected so that the images and FWHM almost matched those of /sup 18/F, which has negligible positron range. However, statistical noise was increased by the deconvolution process and it was not practical to recover the full spatial resolution of the tomograph.<>