Multi-Channel Magnetocardiogardiography System Based on Low-Tc SQUIDs in an Unshielded Environment
Multi-Channel Magnetocardiogardiography System Based on Low-Tc SQUIDs in an Unshielded Environment
复制标题
非屏蔽环境下基于 Low-Tc SQUID 的多通道心磁图系统
DOI:
10.1016/j.phpro.2012.06.161
复制
发表时间:
2012
期刊:
影响因子:
--
通讯作者:
Xiaoming Xie
中科院分区:
文献类型:
--
作者:
X. Kong;Shulin Zhang;Yongliang Wang;J. Zeng;Xiaoming Xie
Magnetocardiography (MCG) using superconducting quantum interference devices (SQUIDs) is a new medical diagnostic tool measuring biomagnetic signals that are generated by the electrical activity of the human heart. This technique is completely passive, contactless, and it has an advantage in the early diagnosis of heart diseases. We developed the first unshielded four-channel MCG system based on low-Tc DC SQUIDs in China. Instead of using a costly magnetically shielded room, the environmental noise suppression was realized by using second-order gradiometers and three-axis reference magnetometer. The measured magnetic field resolution of the system is better than 1 pT, and multi-cycle human heart signals can be recorded directly. Also, with the infrared positioning system, 48 points data collection can be realized by moving the non-magnetic bed nine times.