Preoperative Accuracy of Selective Laser Sintering (SLS) in Craniofacial 3D Modeling: Comparison with Patient CT Data
Preoperative Accuracy of Selective Laser Sintering (SLS) in Craniofacial 3D Modeling: Comparison with Patient CT Data
复制标题
DOI:
10.1055/s-0028-1109204
复制
发表时间:
2009-07-01
影响因子:
1.8
通讯作者:
Hammer, B.
中科院分区:
文献类型:
--
作者:
Kaim, A. H.;Kirsch, E. C.;Hammer, B.
Purpose: To evaluate the accuracy of craniofacial 3D models produced in the standardized selective laser sintering (SLS) technique from multislice. computed tomography (MSCT) data sets in comparison with patient data and to investigate the effect of potential causes of inaccuracies.Materials and Methods: 19 models were considered and examined by MSCT. The patient CT data used for 3D modeling was analyzed and compared to the 3D model data. 15 anatomical landmarks were defined and 20 distances were digitally measured. The digital measurements of both CT data sets were compared to manually measured distances of the SLS model.Results: There was not a statistically significant difference (p