Reliability evaluation of R-R interval measurement status for time domain heart rate variability analysis with wearable ECG devices
Reliability evaluation of R-R interval measurement status for time domain heart rate variability analysis with wearable ECG devices
复制标题
使用可穿戴心电图设备进行时域心率变异性分析的 R-R 间隔测量状态的可靠性评估
DOI:
10.1109/embc.2017.8037072
复制
发表时间:
2017
期刊:
影响因子:
--
通讯作者:
Tomohiro Yamada
中科院分区:
文献类型:
--
作者:
Kana Eguchi;Ryosuke Aoki;Kazuhiro Yoshida;Tomohiro Yamada
Electrocardiograms (ECGs) captured by wearable ECG devices easily contain artifacts because of measurement faults. Since the frequency characteristics of artifacts are quite similar to those of R waves, it may result in R-R interval (RRI) miscalculations. To enable accurate heart rate variability (HRV) analysis in daily life, this paper proposes a method to reliably evaluate RRI measurement status, one that uses the electric potential characteristics of the QRS complex. Initial evaluation results show that it has the potential to distinguish miscalculated RRIs. Also proposed is a new RRI outlier exclusion method that uses the aforementioned method. Time domain measures of HRV derived from the proposed RRI outlier exclusion method are found to be more accurate than those derived from the conventional one.