IMPROVEMENTS OF DNA SEQUENCING GELS
IMPROVEMENTS OF DNA SEQUENCING GELS
复制标题
DOI:
10.1016/0003-2697(81)90031-2
复制
发表时间:
1981-01-01
影响因子:
2.9
通讯作者:
ANSORGE, W
中科院分区:
文献类型:
--
作者:
GAROFF, H;ANSORGE, W
Three simple modifications of DNA sequencing gels are described which all result in improved oligonucleotide resolution as visualized by autoradiography. It was possible to reduce the thickness of the gel to 0.2 mm by using new gel molding techniques. The gel could be dried without any distortions of its dimensions by prior binding of the gel to the surface of the glass plate. A uniform high temperature was obtained in all parts of the gel during electrophoresis by replacing one of the glass plates with an inexpensive thermostating plate with circulating water. The use of this heating plate resulted in a straight band pattern all over the gel and also in the resolution of such bands which were not resolved in other electrophoresis systems.