C-arm calibration--is it really necessary?
C-arm calibration--is it really necessary?
复制标题
C型臂校准——真的有必要吗?
DOI:
10.1007/11566465_79
复制
发表时间:
2005
期刊:
影响因子:
--
通讯作者:
Fichtinger,Gabor
中科院分区:
文献类型:
--
作者:
Jain,Ameet;Kon,Ryan;Zhou,Yu;Fichtinger,Gabor
C-arm fluoroscopy is modelled as a perspective projection, the parameters of which are estimated through a calibration procedure. It has been universally accepted that precise intra-procedural calibration is a prerequisite for accurate quantitative C-arm fluoroscopy guidance. Calibration, however, significantly adds to system complexity, which is a major impediment to clinical practice. We challenge the status quo by questioning the assumption that precise intra-procedural calibration is really necessary. We derived theoretical bounds for the sensitivity of 3D measurements to mis-calibration. Experimental results corroborated the theory in that mis-calibration in the focal spot by as much as 50mmstill allows for tracking with an accuracy of 0.5mmin translation and 0.65oin rotation, and such mis-calibration does not impose any additional error on the reconstruction of small objects.