In vivo thermal ablation monitoring by echo decorrelation imaging
In vivo thermal ablation monitoring by echo decorrelation imaging
复制标题
通过回波去相关成像进行体内热消融监测
DOI:
10.1063/1.4757364
复制
发表时间:
2012
影响因子:
2.3
通讯作者:
T. D. Mast
中科院分区:
文献类型:
--
作者:
S. Subramanian;S. Rudich;Amel Alqadah;Chandra Karunakaran;T. D. Mast
In the studies reported here, the ability of echo decorrelation imaging to predict thermal ablation in vivo is tested. Radiofrequency ablation (RFA) was performed on a normal swine liver (N=5) and ultrasound ablation using image-ablate arrays was performed on a rabbit implanted with VX2 tumors (N=3). Echo decorrelation imaging was able to predict ablation and overtreatment with high accuracy for both RFA and ultrasound ablation. The effect of tissue motion was studied by comparing the mean cumulative decorrelation with and without treatment. The mean cumulative decorrelation with treatment was found to be significantly greater than the mean cumulative decorrelation without treatment (p=0.0347 for RFA, p=0.024 for ultrasound ablation). The accuracy of ablation predictions was tested using receiver operating characteristic curves and the area under the ROC curve (AUROC) was significant for ablation and overtreatment for RFA (N=5). However, the AUROC was significant for only detecting ablation for the ultras...