Influence of Tricuspid Bioprosthetic Mitral Valve Orientation Regarding the Flow Field Inside the Left Ventricle: In Vitro Hydrodynamic Characterization Based on 2D PIV Measurements
Influence of Tricuspid Bioprosthetic Mitral Valve Orientation Regarding the Flow Field Inside the Left Ventricle: In Vitro Hydrodynamic Characterization Based on 2D PIV Measurements
复制标题
DOI:
10.1111/aor.12515
复制
发表时间:
2016-02-01
影响因子:
2.4
通讯作者:
Yanagihara, Jurandir I.
中科院分区:
文献类型:
--
作者:
Bazan, Ovandir;Ortiz, Jayme P.;Yanagihara, Jurandir I.
The flow patterns of a prosthetic heart valve in the aortic or mitral position can change according to its type and orientation. This work describes the use of 2D particle image velocimetry (PIV) applied to the in vitro flow fields characterization inside the upper part of a left ventricular model at various heart rates and as a function of two orientations of stented tricuspid mitral bioprostheses. In the ventricular model, each mitral bioprosthesis (27 and 31 mm diameter) was installed in two orientations, rotated by 180 degrees, while the aortic bileaflet mechanical valve (27 mm diameter) remained in a fixed orientation. The results (N = 50) showed changes in the intraventricular flow fields according to the mitral bioprostheses positioning. Also, changes in the aortic upstream velocity profiles were noticed as a function of mitral orientations.