Automated system for solvent extraction kinetic studies
Automated system for solvent extraction kinetic studies
复制标题
用于溶剂萃取动力学研究的自动化系统
DOI:
10.1021/ac00250a062
复制
发表时间:
1982
影响因子:
7.4
通讯作者:
H. Freiser
中科院分区:
文献类型:
--
作者:
H. Watarai;Larry Cunningham;H. Freiser
CONCLUSIONS We have described our successful modification of HP 5985B, HP 5982, and HP 5995 GC/MS units for direct introduction of fused silica columns in close proximity to the ionizing beam in the ion source of the instrument. For the HP 5985B and HP 5982, the modification increased transmission of com-pounds with long GC retention times. Not only was sensitivity improved for the HP 5995 but, more importantly, thechemical reactivity and selectivity of the isolation valve were eliminated by its removal. In all three cases, we are confident that our approach was the best available solution. Though the man-ufacturer’s designed flexibility of use (packed or capillary columns) and ion source vacuum isolation have been sacrificed, these losses have been inconsequentialin view of substantial gains in data quality and sensitivity.(Sketches of the hard-ware modification are available on request to RAH) ACKNOWLEDGMENT