The large Debye-Scherrer camera installed at SPring-8 BL02B2 for charge density studies
The large Debye-Scherrer camera installed at SPring-8 BL02B2 for charge density studies
复制标题
DOI:
10.1016/s0168-9002(01)00639-8
复制
发表时间:
2001-07-21
影响因子:
1.4
通讯作者:
Ikeda, N
中科院分区:
文献类型:
--
作者:
Nishibori, E;Takata, M;Ikeda, N
The design and performance of a large Debye-Scherrer Camera with imaging plate (IP) as a detector. which was very recently installed at SPring-8, BL02B2, is reported. By taking advantage of high beam quality of SPring-8, the camera enables one a rapid collection of a high counting statistics and high angular resolution powder pattern, which can lead to accurate structure analyses. The camera also provides easy access to structural changes at varied temperatures between 15-1000 K. The camera provides a rapid and accurate powder diffraction system utilizing third generation SR. (C) 2001 Elsevier Science B.V. All rights reserved.