Carotid artery distensibility - A reliability study
Carotid artery distensibility - A reliability study
复制标题
DOI:
10.7863/jum.2007.26.9.1157
复制
发表时间:
2007-09-01
影响因子:
2.3
通讯作者:
Rundek, Tatjana
中科院分区:
文献类型:
--
作者:
Godia, Elisa Cuadrado;Madhok, Rishi;Rundek, Tatjana
Objective. Carotid distensibility (CD) is a measure of carotid artery elasticity that has been introduced as a risk factor for cardiovascular disease. information regarding reproducibility of sonographic CD measures is limited. The objective of this study was to evaluate the inter-reader reliability of sonographic measurements of common carotid artery (CCA) diameters and derived metrics of CD. Methods. Two independent readers (R1 and R2) measured the systolic diameter (SD) and diastolic diameter (DD) for the right CCA from the B/M-mocle sonographic registrations among 118 subjects. The derived CD metrics (strain, elastic modulus [E], stiffness [P], and CID) were calculated. The inter-reader type 3 intraclass correlation coefficients (ICC3,1) for carotid diameters were calculated. Results. The mean SDs standard deviation were 7.15 +/- 1.43 mm for R1 and 7.24 +/- 1.43 mm for R2. The mean DDs were 6.71 +/- 1.36 mm for R1 and 6.68 +/- 1.41 mm for R2. The mean differences of SD and DD between R1 and R2 were 0.08 +/- 0.40 mm (paired t test, P =.04) and 0.03 +/- 0.43 mm (paired t test, P =.46), respectively Inter-reader type 3 intraclass correlation coefficients were 0.96 for SD and 0.95 for DD. We observed a significant association of demographics with carotid diameters but not with derived CD metrics or risk factors. Conclusions. Our results suggest good reproducibility of CCA diameters measured with B/Mmode sonography. However, very small changes in linear measurements of carotid diameters can have big effects on estimates of arterial mechanical properties such as strain and Young's modulus. The standard boundary identification methods may not be precise and reproducible enough for use in a clinical setting.