Preclinical evaluation of Millipede 088 intracranial aspiration catheter in cadaver and in vitro thrombectomy models.

Preclinical evaluation of Millipede 088 intracranial aspiration catheter in cadaver and in vitro thrombectomy models.
复制标题

Millipede 088 颅内抽吸导管在尸体和体外血栓切除模型中的临床前评估。

DOI:
10.1136/neurintsurg-2020-016218
复制
发表时间:
2021-05
影响因子:
4.8
通讯作者:
Nogueira RG
Nogueira RG
中科院分区:
医学1区
文献类型:
--
作者:
Fitzgerald S;Ryan D;Thornton J;Nogueira RG

文献摘要

参考文献

被引文献

相似文献

Larger bore aspiration catheters are expected to significantly improve the speed and completeness of acute stroke revascularization. To evaluate the navigability and clot retrieval performance of a novel 8Fr aspiration catheter, Millipede 088 (Perfuze Ltd), using fresh-frozen cadavers and an in vitro thrombectomy model, respectively. Cadaveric study: Transfemoral catheterization of the intracranial arteries was performed in six cadavers, allowing evaluation of navigation to 12 middle cerebral arteries (MCAs) and six basilar arteries. Commercially available 6Fr aspiration catheters (SOFIA Plus, Microvention) were used as controls. In vitro study: Three human blood clot phenotypes were created; red blood cell-rich, mixed, and fibrin/platelets-rich. Two clot sizes, resulting in occlusion of the internal carotid artery (ICA) and MCA-M1 were investigated. Endpoints were first-pass effect (FPE), first-pass complete ingestion, and second-pass recanalization. Cadaveric study: Both the Millipede 088 and SOFIA Plus devices reached the distal MCA-M1 and the basilar artery in 10/12 and 2/2 of the navigation attempts, respectively. In the two instances of unsuccessful navigation, neither device was able to cross the ophthalmic artery. In vitro study: In 10 mm long M1 occlusions, Millipede 088 achieved 100% FPE versus 40% for 6Fr devices (p>0.001). In 20 mm long ICA occlusions, Millipede 088 achieved 100% removal success within two passes in each clot phenotype compared with an average of 27% for 6Fr devices (p>0.001). Navigation of the Millipede 088 catheter to the MCA-M1 and basilar artery is feasible in a cadaver model. Millipede 088 demonstrates superiority over 6Fr aspiration catheters for three representative clot phenotypes at the most common sites of occlusion in an in vitro vasculature model.
DOI: 10.1161/strokeaha.116.016149
发表时间: 2017-06-01
期刊: STROKE
影响因子: 8.3
作者:
Blanc, Raphael;Redjem, Hocine;Piotin, Michel
通讯作者: Piotin, Michel
DOI: 10.1136/neurintsurg-2019-015215
发表时间: 2020-04-01
影响因子: 4.8
作者:
Ducroux, Celina;Piotin, Michel;Fahed, Robert
通讯作者: Fahed, Robert
DOI: 10.3174/ajnr.a6117
发表时间: 2019-08-01
影响因子: 3.5
作者:
O'Neill, D.;Griffin, E.;Thornton, J.
通讯作者: Thornton, J.
DOI: 10.1111/j.1469-7580.2010.01297.x
发表时间: 2010-11-01
期刊: JOURNAL OF ANATOMY
影响因子: 2.4
作者:
Demertzis, Stefanos;Hurni, Samuel;Van den Berg, Jos
通讯作者: Van den Berg, Jos
DOI: 10.1056/nejmoa1411587
发表时间: 2015-01-01
影响因子: 158.5
作者:
Berkhemer, O. A.;Fransen, P. S. S.;Dippel, D. W. J.
通讯作者: Dippel, D. W. J.