Effect of Environment on Fatigue Failure of Controlled Memory Wire Nickel-Titanium Rotary Instruments
Effect of Environment on Fatigue Failure of Controlled Memory Wire Nickel-Titanium Rotary Instruments
复制标题
DOI:
10.1016/j.joen.2011.12.002
复制
发表时间:
2012-03-01
影响因子:
4.2
通讯作者:
Haapasalo, Markus
中科院分区:
文献类型:
--
作者:
Shen, Ya;Qian, Wei;Haapasalo, Markus
Introduction: This study examined the fatigue behavior of 2 types of nickel-titanium (NiTi) instruments made from a novel controlled memory NiTi wire (CM wire) under various environment conditions. Methods: Three conventional superelastic NiTi instruments of Pro File (Dentsply Maillefer, Ballaigues, Switzerland), Typhoon (Clinician's Choice Dental Products, New Milford, CT), and DS-SS0250425NEYY (Clinician's Choice Dental Products) and 2 new CM wire instruments of Typhoon CM and DS-SS0250425NEYY CM were subjected to rotational bending at the curvature of 35 in air, deionized water, 17% EDTA, or deionized water after immersion in 6% sodium hypochlorite for 25 minutes, and the number of revolutions of fracture (Ni) was recorded. The fracture surface of all fragments was examined by a scanning electron microscope. The crack-initiation sites and the percentage of dimple area to the whole fracture cross-section were noted. Results: Two new CM Wire instruments yielded an improvement of >4 to 9 times in Nf than conventional NiTi files with the same design under various environments (P