A CONTINUOUS, IMPLANTABLE LACTATE SENSOR
A CONTINUOUS, IMPLANTABLE LACTATE SENSOR
复制标题
DOI:
10.1021/ac00105a010
复制
发表时间:
1995-05-01
影响因子:
7.4
通讯作者:
GOUGH, DA
中科院分区:
文献类型:
--
作者:
BAKER, DA;GOUGH, DA
An implantable sensor for continuous monitoring of lactate in the bloodstream or tissues has been made. The sensor includes a lactate electrode based on immobilized L-lactate oxidase coupled to a potentiostatic oxygen electrode and an oxygen reference electrode to account for local oxygen. This design renders the response essentially independent of oxygen concentration. In vitro characterization demonstrated that the sensor can respond specifically to lactate over a broad concentration range. This response is stable for more than 1 week curing continuous in vitro operation at body temperature. As an acute implant in the canine right atrium, the sensor produced a continuous recording of lactate that was in good agreement with conventional, discrete blood lactate assays.