High-speed bubble-segmented blood sampler for the rat.
High-speed bubble-segmented blood sampler for the rat.
复制标题
大鼠高速分段气泡血液采样器。
DOI:
10.1152/jappl.1984.57.4.1284
复制
发表时间:
1984
期刊:
影响因子:
--
通讯作者:
G. Abbott
中科院分区:
文献类型:
--
作者:
M. Graham;G. Abbott
An arterial blood sampler for use in the conscious rat is described. With this apparatus it is possible to obtain small (10 microliter) whole-blood or plasma samples as frequently as 1/s and to derive accurate arterial time-concentration curves in the first 60-120 s after compounds are injected for regional blood flow or pharmacokinetic measurements. The blood is withdrawn from an implanted arterial catheter into polyethylene tubing at a constant rate while bubbles are introduced at regular intervals via a side channel into the column of blood. Although some dispersion of blood samples occurs as the tubing is traversed, this can be mathematically corrected. However, correction is unnecessary if the information of interest is the area under the time-concentration curve.