RapidArc treatment verification in 3D using polymer gel dosimetry and Monte Carlo simulation

RapidArc treatment verification in 3D using polymer gel dosimetry and Monte Carlo simulation
复制标题

DOI:
10.1088/0031-9155/55/17/001
复制
发表时间:
2010-09-07
影响因子:
3.5
通讯作者:
Back, Sven A. J.
Back, Sven A. J.
中科院分区:
工程技术2区
文献类型:
--
作者:
Ceberg, Sofie;Gagne, Isabelle;Back, Sven A. J.

文献摘要

被引文献

相似文献

本研究的目的是使用 3D 凝胶测量和蒙特卡罗 (MC) 模拟验证体积弧治疗技术 (RapidArc (TM)) 产生的先进不均匀剂量分布。将 TPS(治疗计划系统)计算的剂量分布与凝胶测量和 MC 模拟进行比较,从而调查计划剂量输送与实际输送之间的任何差异。此外,使用重复的凝胶测量来研究递送的再现性。使用单弧旋转和 3.3 Gy 的目标剂量将前列腺治疗计划传送到 1.3 升 nPAG 凝胶体模。使用 1.5 T 扫描仪对凝胶进行磁共振成像。使用 VIMC-Arc 代码计算 MC 剂量分布。对于 TPS 与凝胶以及 TPS 与 MC,在 90% 等剂量表面 (VOI90) 封闭的体积内逐体素计算相对吸收剂量差异。以相同的方式研究验证方法、MC 与凝胶之间以及两次重复凝胶测量之间的差异。对于所有体积比较,平均值在 1% 以内,差异的标准偏差在 2.5% (1SD) 以内。分别使用 50% 等剂量表面 (VOI50) 和 90% 等剂量表面 (VOI90) 包围的体积内使用 3%/3 mm 和 5%/5 mm(% 剂量差和 mm 一致距离)伽马标准,对剂量矩阵之间进行 3D 伽马分析。所有比较都得出了非常高的通过率。在 VOI50 和 VOI90 内,超过 95% 的 TPS 点在凝胶测量和 MC 模拟的 3%/3 mm 范围内。此外,重复的凝胶测量显示出出色的一致性,表明输送具有可重复性。通过 MC 模拟和凝胶测量,该验证研究成功证明 RapidArc (TM) 计划能够准确计算并按计划交付。
The aim of this study was to verify the advanced inhomogeneous dose distribution produced by a volumetric arc therapy technique (RapidArc (TM)) using 3D gel measurements and Monte Carlo (MC) simulations. The TPS (treatment planning system)-calculated dose distribution was compared with gel measurements and MC simulations, thus investigating any discrepancy between the planned dose delivery and the actual delivery. Additionally, the reproducibility of the delivery was investigated using repeated gel measurements. A prostate treatment plan was delivered to a 1.3 liter nPAG gel phantom using one single arc rotation and a target dose of 3.3 Gy. Magnetic resonance imaging of the gel was carried out using a 1.5 T scanner. The MC dose distributions were calculated using the VIMC-Arc code. The relative absorbed dose differences were calculated voxel-by-voxel, within the volume enclosed by the 90% isodose surface (VOI90), for the TPS versus gel and TPS versus MC. The differences between the verification methods, MC versus gel, and between two repeated gel measurements were investigated in the same way. For all volume comparisons, the mean value was within 1% and the standard deviation of the differences was within 2.5% (1SD). A 3D gamma analysis between the dose matrices were carried out using gamma criteria 3%/3 mm and 5%/5 mm (% dose difference and mm distance to agreement) within the volume enclosed by the 50% isodose surface (VOI50) and the 90% isodose surface (VOI90), respectively. All comparisons resulted in very high pass rates. More than 95% of the TPS points were within 3%/3 mm of both the gel measurement and MC simulation, both inside VOI50 and VOI90.Additionally, the repeated gel measurements showed excellent consistency, indicating reproducible delivery. Using MC simulations and gel measurements, this verification study successfully demonstrated that the RapidArc (TM) plan was both accurately calculated and delivered as planned.